Targeted Intraoperative Radiotherapy Is a Safe Approach for Patients with Pacemakers: A Case Study and Literature Review

靶向术中放射治疗对装有心脏起搏器的患者来说是一种安全的方法:病例研究和文献综述

阅读:1

Abstract

Case reports detailing the effects of targeted intraoperative radiation therapy (IORT) on patients with cardiac pacemakers (PMs) are rare. This growing population sub-group requiring IORT and lack of standardized guidelines necessitate more practical published research. An 81-year-old patient with clinical stage II, T1 N0 grade III, triple-negative invasive ductal carcinoma and an implanted single-lead chamber PM (VVIR mode, model: Biotronik, type Effecta SR) received targeted intraoperative radiotherapy at the time of wide local excision and sentinel lymph node biopsy. It presents the shortest distance between the outer diameter of the PM and IORT applicator in literature. Target IORT was performed utilizing an Intrabeam device (50 kV, Carl Zeiss Surgical, Oberkochen, Germany). This case elucidates the successful use of targeted IORT for breast-conserving surgery in a patient with a single ipsilateral chamber VVIR mode PM. No device failure or malfunction was reported for the PM before, during, or after the procedure. These findings support the use of targeted IORT for patients diagnosed with early-stage breast carcinomas who have a PM implanted. However, further research is needed to understand the safety of other methods and devices for IORT patients with cardiac implantable electronic devices.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。