Vault Height Is a Key Predictive Factor for Anterior Segment Measurement Error by IOLMaster 700 in Eyes With Phakic Intraocular Lens

对于有晶体眼人工晶状体植入的患者,拱高是 IOLMaster 700 前节测量误差的关键预测因素。

阅读:2

Abstract

PURPOSE: To identify risk factors of ocular anterior segment measurement error by the IOLMaster 700 in eyes implanted with an implantable Collamer lens (ICL). METHODS: In total, 152 patients with clear lens (152 eyes, group 1) and another 32 cataract patients (57 eyes, group 2) who underwent ICL implantation were included, and the presence of measurement error by the IOLMaster 700 was determined based on B-scan images. The risk factors for measurement error were evaluated by logistic regression, and the optimal threshold was determined using receiver operating characteristic analysis. RESULTS: The ICL was misidentified as the anterior surface of the crystalline lens in 51.97% of eyes (79/152) in group 1 and 80.70% of eyes (46/57) in group 2. For every 100-µm decrease in the vault height, a 3.57- and 5.78-fold increase in the risk of measurement error was observed in group 1 and group 2, respectively. We identified an optimal threshold of the vault height at 389.47 µm for predicting biometric measurement error in eyes implanted with ICL, which showed an area under the curve of 0.93 (95% confidence interval, 0.90-0.97), a sensitivity of 0.87, and a specificity of 0.86. CONCLUSIONS: Patients with ICL implantation, particularly those with a vault height less than 389.47 µm, are at a greater risk of anterior segment biometric measurement error by the IOLMaster 700. TRANSLATIONAL RELEVANCE: The threshold of vault height can help to identify high-risk patients and further optimize biometric measurement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。