Comparison of the Third Wave Invader human papillomavirus (HPV) assay and the digene HPV hybrid capture 2 assay for detection of high-risk HPV DNA

第三次入侵人类乳头瘤病毒 (HPV) 检测与 digene HPV 杂交捕获 2 检测在检测高危 HPV DNA 方面的比较

阅读:9
作者:C C Ginocchio, D Barth, F Zhang

Abstract

This study compared the clinical performance of the Digene Hybrid Capture 2 (HC2) assay to that of a prototype Third Wave Invader human papillomavirus (HPV) (IHPV) analyte-specific reagent-based assay for the detection of oncogenic or "high-risk" (HR) HPV DNA using liquid-based cytology specimens. In total, 821 ThinPrep vials were tested using both assays. In accordance with the type-specific probes contained within each test, the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) for the IHPV assay were 95.9%, 97.6%, 97.5%, and 96.1%, respectively, and those for the HC2 assay were 98.1%, 86.2%, 87.1%, and 97.9%. Overall, the sensitivity and NPV were comparable between the assays, but the IHPV assay demonstrated a better specificity and PPV, since the IHPV assay had fewer false-positive HR HPV results. The incorporation of an internal control to evaluate the cellularity of the test material is an important feature of the IHPV assay and should reduce the risk of false-negative results due to insufficient sample collection rather than the lack of HR HPV DNA. An additional benefit of the IHPV assay was the smaller sample volume required (1 ml versus 4 ml for the HC2 assay).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。