Changes in preoperative corneal measurements following same-day intraocular pressure testing with rebound tonometry

术前角膜测量值在当日使用回弹式眼压计进行眼压测试后的变化

阅读:1

Abstract

PURPOSE: To evaluate the extent to which rebound tonometry affects corneal surface properties and preoperative corneal measurements. SETTING: Four cornea specialty private practices. DESIGN: Prospective case series. METHODS: Visual acuity testing, corneal topography, keratometry, and grading of corneal staining were performed on both eyes of 60 randomly selected, previously scheduled patients. Technicians then performed rebound tonometry on one randomly selected eye only. Immediately following, intraocular pressure measurement, corneal topography, keratometry, and corneal staining were repeated on both eyes. RESULTS: None of the 60 study eyes developed increased staining scores following intraocular pressure testing with the Icare ic100. For corneal staining, mean keratometry, and total corneal cylinder, no statistically significant difference was found from the first measurement to the second measurement between the study eyes and control eyes. CONCLUSION: Rebound tonometry with the Icare ic100 may be used on any patient at any time during the exam without affecting the results of other tests, allowing clinicians to test intraocular pressure prior to preoperative cataract or refractive surgery measurements on the same day. This may allow for significant improvement in patient flow in the office and save patients from the cost and time of extra visits.