Abstract
BACKGROUND: Leadless endocardial left ventricular (LV) pacing resynchronization therapy is a novel solution for patients with heart failure (HF) in whom conventional cardiac resynchronization therapy (CRT) failed. METHODS: PubMed and the Cochrane Library were searched for relevant cohort studies. Clinical outcomes of interest such as ejection fraction (EF), QRS duration (QRSd), and left ventricular end-systolic volume (LVESV) were extracted and analyzed. RESULTS: Five studies involving 175 HF patients for WiSE CRT were included, and patients were followed-up for 6 months. The implanted success rate ranged from 76.5 to 100%. WiSE CRT resulted in significantly narrower QRSd [mean difference (MD): -38.21 ms, 95% confidence interval (CI): -44.36 to -32.07, p < 0.001], improved left ventricular ejection fraction (MD: 6.07%, 95% CI: 4.43 to 7.71, I(2) = 0%, p < 0.001), reduced left ventricular end-systolic volume (MD: -23.47 ml, 95% CI: -37.18 to -9.13, p < 0.001), and reduced left ventricular end-diastolic volume (MD: -24.02 ml, 95% CI: -37.01 to -11.03, p = 0.02). CONCLUSION: Evidence from current studies suggests that leadless endocardial LV pacing resynchronization is effective for HF patients who failed conventional CRT or needed a device upgrade, and it may be an interesting rescue therapy.