Linearization improves the repeatability of quantitative dynamic contrast-enhanced MRI

线性化提高了定量动态对比增强磁共振成像的重复性

阅读:1

Abstract

PURPOSE: The purpose of this study was to compare the repeatabilities of the linear and nonlinear Tofts and reference region models (RRM) for dynamic contrast-enhanced MRI (DCE-MRI). MATERIALS AND METHODS: Simulated and experimental DCE-MRI data from 12 rats with a flank tumor of C6 glioma acquired over three consecutive days were analyzed using four quantitative and semi-quantitative DCE-MRI metrics. The quantitative methods used were: 1) linear Tofts model (LTM), 2) non-linear Tofts model (NTM), 3) linear RRM (LRRM), and 4) non-linear RRM (NRRM). The following semi-quantitative metrics were used: 1) maximum enhancement ratio (MER), 2) time to peak (TTP), 3) initial area under the curve (iauc64), and 4) slope. LTM and NTM were used to estimate K(trans), while LRRM and NRRM were used to estimate K(trans) relative to muscle (R(Ktrans)). Repeatability was assessed by calculating the within-subject coefficient of variation (wSCV) and the percent intra-subject variation (iSV) determined with the Gage R&R analysis. RESULTS: The iSV for R(Ktrans) using LRRM was two-fold lower compared to NRRM at all simulated and experimental conditions. A similar trend was observed for the Tofts model, where LTM was at least 50% more repeatable than the NTM under all experimental and simulated conditions. The semi-quantitative metrics iauc64 and MER were as equally repeatable as K(trans) and R(Ktrans) estimated by LTM and LRRM respectively. The iSV for iauc64 and MER were significantly lower than the iSV for slope and TTP. CONCLUSION: In simulations and experimental results, linearization improves the repeatability of quantitative DCE-MRI by at least 30%, making it as repeatable as semi-quantitative metrics.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。