Shearwave Elastography Increases Diagnostic Accuracy in Characterization of Breast Lesions

剪切波弹性成像技术提高了乳腺病变鉴别诊断的准确性

阅读:1

Abstract

The purpose of this study was to investigate the diagnostic efficacy of shearwave elastography (SWE) in differentiating between benign and malignant breast lesions.One hundred and fifty-nine lesions were assessed using B-mode ultrasound (US) and SWE parameters were recorded (Emax, Emean, Emin, Eratio, SD). SWE measurements were then correlated with histopathological diagnosis.The final sample contained 85 benign and 74 malignant lesions. The maximum stiffness (Emax) with a cutoff point of ≥ 56.0 kPa (based on ROC curves) provided sensitivity of 100.0%, specificity of 97.6%, positive predictive value (PPV) of 97.4%, and negative predictive value (NPV) of 100% in detecting malignant lesions. A cutoff of ≥80 kPa managed to downgrade 95.5% of the Breast Imaging-Reporting and Data System (BI-RADS) 4a lesions to BI-RADS 3, negating the need for biopsy. Using a combination of BI-RADS and SWE, the authors managed to improve the PPV from 2.3% to 50% in BI-RADS 4a lesions.SWE of the breast provides highly specific and sensitive quantitative values that are beneficial in the characterization of breast lesions. Our results showed that Emax is the most accurate value for differentiating benign from malignant lesions.