Phantom validation of a novel RSA-based impingement metric to assess component-on-component impingement risk

利用体模验证一种基于RSA的新型碰撞指标,以评估组件间的碰撞风险

阅读:1

Abstract

Component-on-component impingement in total hip arthroplasty may lead to post-operative complications including dislocation. Despite numerous clinical studies focusing on reducing this risk, assessment methods remain limited to qualitative radiography, finite element analysis, and cadaver studies. There is a need for more precise measurements of impingement in the research setting. We aimed to validate a novel RSA-based impingement metric to measure component-on-component impingement in vivo. A phantom experiment of a standard metal-on-polyethylene total hip system was performed. RSA examinations were performed as typical for a traditional weight-bearing RSA exam for large joints. The phantom was placed in 10 possible impinged positions and one neutral position. Double exposure radiographs were taken to measure repeatability. The closest distance between the skirt of the head and the inner circumference of the acetabular cup liner was measured to assess impingement risk. Distances between the closest point of the hood to the edge of the cup in 10 impinged positions ranged from 0.05 to 1.03 mm, with the average being 0.67 mm. In the neutral position, the distance measured is 11.02 mm. Excellent repeatability was observed, with a standard deviation of 0.03 mm with an r value of 0.09. A validated RSA-based risk metric was established to evaluate in vivo hip impingement. A 1 mm threshold may be proposed to define impingement where distances approaching 1.00 mm are at a greater risk of impingement. This simplified metric holds promise for upcoming clinical studies on component-on-component impingement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。