Optimization and validation of patient-based real-time quality control procedure using moving average and average of normals with multi-rules for TT3, TT4, FT3, FT3, and TSH on three analyzers

在三种分析仪上,采用移动平均法和正常值平均值结合多规则,对基于患者的实时质量控制程序进行优化和验证,用于检测 TT3、TT4、FT3、FT3 和 TSH 水平。

阅读:1

Abstract

BACKGROUND: We have designed a patient-based real-time quality control (PBRTQC) procedure to detect analytical shifts and review analytical trends of measurement procedures. METHODS: All the nine months' patient results of total thyroxine (TT4), total triiodothyronine (TT3), free thyroxine (FT4), free triiodothyronine (FT3), and thyrotropin (TSH) measured by three identical analyzers were divided into three groups according to the source of inpatient patients, outpatient patients, and healthy people. The data in each group were truncated by optimized Box-Plot method and normalized by Box-Cox method if necessary. The z-score charts of internal quality control (IQC) samples' results and PBRTQC data were drawn by IQC levels and groups, respectively. The analytical shifts and analytical trends were detected by multi-rules of 2-2S rules and moving average rules. The performances of PBRTQC were compared with the BIQC in which IQC samples were measurand only once per day at the beginning of the analytical batch. Twelve quality control cases were listed to validate the performances. RESULTS: All the five analytes presented normal distributions when the parameter n of Box-Plot method was 1.2. The percentages of excluded data ranged from 2.9% to 11.6%. 31 and 14 rejections triggered in PBRTQC and BIQC, respectively. 96.8% of the shift rejections in PBRTQC were trend-related shifts and calibration-related shifts, while the proportion was 85.7% in BIQC but 78.6% of the shift rejections in TSH. 25.7% and 8.6% of 105 calibration events which caused analytical shifts were detected by PBRTQC and BIQC, respectively. However, the performance of PBRTQC was not well in TSH because of its large coefficient of variation. CONCLUSIONS: The optimized PBRTQC is high efficiency than BIQC in detecting analytical shifts, trends, and calibration events. The PBRTQC can be used as a low-cost supplementary procedure to IQC every day, especially at the end of the analytical batch on that day when the within-individual biological variation of analyte is not larger than its coefficient of variation in IQC. Further optimization and validation of PBRTQC are still needed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。