Analytical Validation of the IMMULITE(®) 2000 XPi Progesterone Assay for Quantitative Analysis in Ovine Serum

IMMULITE(®) 2000 XPi 孕酮测定法在绵羊血清定量分析中的分析验证

阅读:1

Abstract

Monitoring circulating progesterone (P4) concentration is an important component of basic and applied reproduction research and clinical settings. IMMULITE® 2000 XPi (Siemens) is a newly upgraded fully automated immunoassay system marketed for human use to measure concentrations of different analytes including P4. Our objective was therefore to characterize the analytical performance of the IMMULITE® 2000 XPi P4 immunoassay across the reportable range in ovine serum. This validation of analytical performance included determining (1) linearity, (2) precision through within-run, and between-run coefficient of variation (CV) calculations, (3) accuracy through bias calculations for spiking-recovery bias and interlaboratory (range and average based) bias for two laboratories across the reportable range (0.2−40 ng/mL). The average within-run and between-run precision (CV%) across the reportable range of the IMMULITE® 2000 XPi P4 immunoassay for serum P4 concentration were both <5%, ranging between 2−8%. The average Observed Total Analytic Error (TEo) reported here for serum P4 concentration across the reportable range was ~30%, ranging from 14.8−59.4%, regardless of the considered bias. Based on these data we conclude that the automated IMMULITE® 2000 XPi P4 immunoassay provides a precise, accurate, reliable, and safe method for measuring P4 concentration ovine serum.