Background
Endothelial injury is an early and enduring feature of cardiovascular disease. Inflammation and hypoxia may be responsible for this, and are often associated with the up-regulation of several transcriptional factors that include Hypoxia Inducible Factor-1 (HIF-1). Although it has been reported that HIF-1α is detectable in plasma, it is known to be unstable. Our
Conclusion
HIF-1α can be measured in plasma, but this requires careful sample collection and storage. The carotid artery balloon injury model is associated with the transient release of HIF-1α into the circulation that probably reflects the hypoxia induced in the artery wall.
Methods
An ELISA for the measurement of HIF-1α in cell-culture medium and plasma was optimized, and the assay was used to determine the best conditions for sample collection and storage. The
Results
The HIF-1α ELISA had a limit of detection of 2.7 pg/mL and was linear up to 1000 pg/ mL. Between and within-assay, the coefficient of variation values were less than 15%. HIF-1α was unstable in cell lysates and plasma, and it was necessary to add a protease inhibitor immediately after collection, and to store samples at -80 °C prior to analysis. The dynamics of HIF-1α release were different for the in vitro and in vivo models. In vitro, HIF-1α reached maximum concentrations approximately 2 h post injury, whereas peak values in plasma and tissues occurred approximately 2 days post injury, in the balloon injury model.