In vitro and in vivo repeatability of abdominal diffusion-weighted MRI

腹部弥散加权磁共振成像的体外和体内重复性

阅读:2

Abstract

OBJECTIVE: To study the in vitro and in vivo (abdomen) variability of apparent diffusion coefficient (ADC) measurements at 1.5 T using a free-breathing multislice diffusion-weighted (DW) MRI sequence. METHODS: DW MRI images were obtained using a multislice spin-echo echo-planar imaging sequence with b-values=0, 100, 200, 500, 750 and 1000 s mm(-2). A flood-field phantom was imaged at regular intervals over 100 days, and 10 times on the same day on 2 occasions. 10 healthy volunteers were imaged on two separate occasions. Mono-exponential ADC maps were fitted excluding b=0. Paired analysis was carried out on the liver, spleen, kidney and gallbladder using multiple regions of interest (ROIs) and volumes of interest (VOIs). RESULTS: The in vitro coefficient of variation was 1.3% over 100 days, and 0.5% and 1.0% for both the daily experiments. In vivo, there was no statistical difference in the group mean ADC value between visits for any organ. Using ROIs, the coefficient of reproducibility was 20.0% for the kidney, 21.0% for the gallbladder, 24.7% for the liver and 28.0% for the spleen. For VOIs, values fall to 7.7%, 6.4%, 8.6% and 9.6%, respectively. CONCLUSION: Good in vitro repeatability of ADC measurements provided a sound basis for in vivo measurement. In vivo variability is higher and when considering single measurements in the abdomen as a whole, only changes in ADC value greater than 23.1% would be statistically significant using a two-dimensional ROI. This value is substantially lower (7.9%) if large three-dimensional VOIs are considered.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。