Radiation doses received in the United Kingdom breast screening programme in 2010 to 2012

2010年至2012年英国乳腺癌筛查项目中受试者接受的辐射剂量

阅读:2

Abstract

OBJECTIVES: To review the radiation doses received by women attending the UK breast-screening programme between 2010 and 2012. To compare doses with previous years and to quantify the impact on dose of changing from analogue to digital imaging and to analyse doses by type of imaging system. METHODS: Measurements of doses to samples of about 50-100 women attending for screening were collected across the whole of the UK breast-screening programme. RESULTS: Data were collected for 87,122 exposures, using 449 X-ray sets, for 25,408 women. The average mean glandular dose (MGD) was 1.79 mGy for mediolateral oblique images and 1.58 mGy for craniocaudal images. The average MGD per two-view examination was 4.01 mGy for film-screen imaging and 3.03 mGy for direct digital radiography (DR) and 4.69 mGy for computed radiography. CONCLUSION: The MGD to women attending breast screening has been reduced on average by about 25% where DR systems have replaced film-screen systems. The dose reduction was greatest for breasts with the largest compressed thickness. There are large variations in dose between the different models of DR system provided by different manufacturers. There should be further work to ensure that all DR systems are operated at the optimal dose level to ensure the best cancer detection while balancing the detriment caused by using radiation. ADVANCES IN KNOWLEDGE: Changes in the radiation dose in breast screening over time have been determined. Specifically, the impact on radiation dose of introducing different types of DR and computed radiography system into breast screening has been quantified.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。