Performance evaluation of the Panther Fusion® respiratory tract panel

Panther Fusion® 呼吸道面板的性能评估

阅读:7
作者:Jolanda J C Voermans, Daphne G J C Mulders, Suzan D Pas, Marion P G Koopmans, Annemiek A van der Eijk, Richard Molenkamp

Background

Respiratory tract infections are among the most common infections during winter season. Rapid diagnostics is required for clinical decision making regarding isolation of patients and appropriate therapy. Objectives: The

Conclusions

The Panther Fusion® provides a random-access system with continuous loading and much shorter sample-to-answer times compared to LDT, albeit with a slightly less clinical sensitivity compared to the LDT.

Results

Analytical sensitivity was generally slightly lower for the Panther Fusion® assays. Clinical specificity and sensitivity was between 96 %-100 % and 71.9 %-100 %, respectively. Discrepant results were found in 146 samples of which 88 samples tested LDT positive / Panther Fusion® negative and 58 samples were LDT negative / Panther Fusion® positive. A total of ten discrepant samples with Ct-values <30 were sequenced to confirm the presence of 7 RV-C not-detected by LDT and 1 RV-A and 2 ADV-2 not detected by Panther Fusion®. Conclusions: The Panther Fusion® provides a random-access system with continuous loading and much shorter sample-to-answer times compared to LDT, albeit with a slightly less clinical sensitivity compared to the LDT.