Total plasma homocysteine measurement: Evaluation of the Abbott immunoassay, comparison with the JEOL ion exchange chromatography and investigation of its clinical utility

总血浆同型半胱氨酸测量:雅培免疫测定法的评估、与 JEOL 离子交换色谱法的比较及其临床实用性的研究

阅读:11
作者:Kai Fung Kevin Suen, Graham R Lee, Michelle Finnegan, Kieran Halton, Ingrid Borovickova, Caoilfionn Trench, Maria C Fitzgibbon

Conclusions

The immunoassay demonstrated robust performance in its verification and showed good comparability with the IEC but with some biases so caution is needed if both are used interchangeably. The immunoassay offers an automated alternative to IEC in the assessment of hyperhomocysteinaemia.

Methods

Following immunoassay method verification, plasma samples of 91 patients were analysed for tHCY using immunoassay and IEC.

Results

For the Abbott immunoassay, accuracy was assessed, with UK NEQAS EQA specimens, by the correlation of our Abbott immunoassay measurements to the Abbott ARCHITECT immunoassay mean (bias = 1.6%), and to the overall immunoassay mean (bias = 2.0%). The total imprecision was 2.7% (11.00 μmol/L), 2.4% (16.80 μmol/L) and 2.8% (24.30 μmol/L) respectively. Bias in linearity assessment was 0.12%-2.58%. The inter-method correlation was strong in Passing-Bablok regression: immunoassay = IEC x0.857 + 2.445 (95% CI: slope = [0.742,0.947], intercept = [1.340,3.582]), with Spearman correlation = 0.803 (p < 0.001). The Bland-Altman plot showed an average difference of -0.284 μmol/L (95% CI: [-1.043,0.474]) with limits of agreement (mean ± 1.96SD) from -7.425 μmol/L to 6.857 μmol/L.No significant difference in tHCY was found using both methods in patients with cerebrovascular diseases and cardiovascular diseases. Most tHCY measurements were within the reference ranges of both methods. All homocystinuria patients had tHCY values above the reference ranges of both methods. Conclusions: The immunoassay demonstrated robust performance in its verification and showed good comparability with the IEC but with some biases so caution is needed if both are used interchangeably. The immunoassay offers an automated alternative to IEC in the assessment of hyperhomocysteinaemia.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。