Abstract
An interlaboratory study was conducted to evaluate a method for the determination of aristolochic acid I, also known as aristolochic acid A, at levels > 2.00 microg/g in botanical species and dietary supplements potentially contaminated with aristolochic acid I. Aristolochic acid I was extracted from various matrixes with aqueous acetonitrile. The amount of aristolochic acid I present was determined by liquid chromatography (LC) using an ultraviolet (UV) detector with confirmation by LC/mass spectrometry (MS). Thirteen blind duplicates were successfully analyzed by 10 collaborators, and aristolochic acid I was successfully confirmed in 1 blind duplicate by 8 collaborators. For repeatability, the relative standard deviation (RSD(r)) ranged from 1.72 to 16.3% and for reproducibility, the RSDR ranged from 5.42 to 19.8%. HorRat values were not applicable for 2 materials but varied from 0.7 to 1.8 for 11 materials. Each collaborating laboratory had calibration curves with correlation coefficients > 0.998. In addition, all of the collaborators that conducted the confirmation were able to verify the identity of aristolochic acid I using LC/MS/MS (using either ion trap or triple quad).