Detection and quantitation of host cell proteins in monoclonal antibody drug products using automated sample preparation and data-independent acquisition LC-MS/MS

使用自动化样品制备和数据独立采集 LC-MS/MS 检测和定量单克隆抗体药物产品中的宿主细胞蛋白

阅读:19

Abstract

Ensuring the removal of host cell proteins (HCPs) during downstream processing of recombinant proteins such as monoclonal antibodies (mAbs) remains a challenge. Since residual HCPs might affect product stability or safety, constant monitoring is required to demonstrate their removal to be below the regulatory accepted level of 100 ng/mg. The current standard analytical approach for this procedure is based on ELISA; however, this approach only measures the overall HCP content. Therefore, the use of orthogonal methods, such as liquid chromatography-mass spectrometry (LC-MS), has been established, as it facilitates the quantitation of total HCPs as well as the identification and quantitation of the individual HCPs present. In the present study, a workflow for HCP detection and quantitation using an automated magnetic bead-based sample preparation, in combination with a data-independent acquisition (DIA) LC-MS analysis, was established. Employing the same instrumental setup commonly used for peptide mapping analysis of mAbs allows for its quick and easy implementation into pre-existing workflows, avoiding the need for dedicated instrumentation or personnel. Thereby, quantitation of HCPs over a broad dynamic range was enabled to allow monitoring of problematic HCPs or to track changes upon altered bioprocessing conditions.