Investigation of morphological changes for the discrimination of nucleated red blood cells and other leukocytes in Sysmex XN hematology analyzer scattergrams using transmission electron microscopy

使用透射电子显微镜研究 Sysmex XN 血液分析仪散点图中有核红细胞和其他白细胞的形态变化

阅读:4
作者:Masako Kaido, Yuri Takagi, Mari Kono, Fumie Nakazawa, Shiori Yamamoto, Atsushi Wada, Takashi Morikawa

Background

The WNR channel of the XN-Series automated hematology analyzer (Sysmex) counts white blood cells (WBCs) and simultaneously performs a differential counting of basophils and nucleated red blood cells (NRBCs). The detection process involves exposing the cells to WNR-specific reagents containing an acidic detergent and a fluorescent dye and measuring the intensity of the forward scattered light (FSC) and side fluorescence light (SFL). Method: We treated isolated peripheral WBCs and NRBCs with specific reagents and assessed the morphological changes in NRBCs and each leukocyte type using transmission electron microscopy (TEM).

Conclusion

We confirmed that the WNR channel differentiates the blood cells on the WNR scattergram based on differences in the amount of residual cytoplasm and nucleic acids.

Results

The results from a flow cytometer (FCM) showed that, after exposure to the reagents, basophils appeared on the highest FSC and SFL areas compared to other leukocytes on the WNR scattergram. Owing to the hemolysis of reticulocytes and erythrocytes, NRBCs that survived the reagent treatment could be distinguished by their lower intensity than those of the other leukocytes on the WNR scattergram. We investigated the significance of the relationship between the TEM and FCM results after the reagent treatment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。