Risk of leakage with a new detachable multi-hard-port containment system for power morcellation during gynecologic laparoscopy: An in vitro study

妇科腹腔镜检查期间使用新型可拆卸多硬端口封闭系统进行电动碎切术的泄漏风险:一项体外研究

阅读:14
作者:Fang Zhao, Wenhui Wang, Bin Ling, Jing Liang

Background

Laparoscopic surgery has been a milestone for minimally invasive surgeries. But safe removal of large uterine tissue is a challenge for minimally invasive procedures, and there still exists concern about the dissemination of benign or occult malignant uterine tissue during the use of the morcellator. Different tissue containment systems have been used in laparoscopic power morcellation. However, a risk of leakage still exists in clinical practice. In this study, we aimed to evaluate leakage and tissue dissemination associated with a new detachable multi-hard-port containment system for tissue removal during laparoscopic myomectomy morcellation.

Conclusions

This study quantified the low leakage rate during morcellation and the improved convenience of operation provided by a new tissue containment system.

Methods

Beef tongue specimens were stained with methylene blue solution and morcellated in a plastic trainer box under laparoscopic guidance. The morcellation test in vitro conditions comprised two different containment systems to simulate laparoscopic power morcellation, specifically a polyurethane bag with two pipes (control group) and a detachable multi-hard-port containment system (experimental group). Insufflation pressure was set at 14 mmHg. Three methods are used to detect the leakage The procedure times were recorded. Thirty trials were performed using a multi-port approach and the two tissue containment systems.

Results

The leakage rate was 0.03% (n = 30) for the experimental group and 26.6% (n = 30) for the control group (p < 0.005). Morcellation time was significantly shorter in the experimental group than in the control group (p < 0.001). Median bag introduction time was shorter in the experimental group than in the control group; however, removal time differences were not significant. Conclusions: This study quantified the low leakage rate during morcellation and the improved convenience of operation provided by a new tissue containment system.