Improved high-force magnetic tweezers for stretching and refolding of proteins and short DNA

改进的高强度磁镊子,用于拉伸和重新折叠蛋白质和短 DNA

阅读:24
作者:Hu Chen, Hongxia Fu, Xiaoying Zhu, Peiwen Cong, Fumihiko Nakamura, Jie Yan

Abstract

Although magnetic tweezers have many unique advantages in terms of specificity, throughput, and force stability, this tool has had limited application on short tethers because accurate measurement of force has been difficult for short tethers under large tension. Here, we report a method that allows us to apply magnetic tweezers to stretch short biomolecules with accurate force calibration over a wide range of up to 100 pN. We demonstrate the use of the method by overstretching of a short DNA and unfolding/refolding a protein of filamin A immunoglobulin domains 1-8. Other potential applications of this method are also discussed.