Left Ventricular Assist Device Inflow Cannula Position May Contribute to the Development of HeartMate II Left Ventricular Assist Device Pump Thrombosis

左心室辅助装置入流导管位置可能导致HeartMate II左心室辅助装置泵血栓形成。

阅读:1

Abstract

BACKGROUND: Pump thrombosis (PT) is a dreaded complication after left ventricular assist device (LVAD) implantation. Problems with inflow cannula (IC) position may precipitate thrombus development. We sought to determine if IC position contributes to the development of PT. METHODS: We conducted a retrospective review of 76 HeartMate II LVAD implants. The angle of the IC (AIC) to the horizontal plane was measured on chest x-rays. Patients who developed PT (PT group) were compared to the remaining patients (control group). RESULTS: The mean age at implantation was 56 ± 14 years, and 82% of the patients were male. Ten patients (13%) developed PT. Six (60%) required device exchange, and 4 (40%) were managed with anticoagulation and/or thrombolysis. The median AIC for all patients at implantation was 59° (range, 38°-98°; 25th-75th interquartile range, 50°-75°). In the PT group, the median AIC was larger at the time of PT diagnosis compared to implantation (70° vs 60°, P = 0.005). In the control group, the median AIC was also larger at follow-up compared to implantation (61° vs 58°, P < 0.001) although to a lesser degree than in the PT group. No difference was seen in the median AIC between the PT group and the control group at implantation (60° vs 58°, respectively; P = 0.668) or at follow-up (70° vs 61°, respectively; P = 0.309). However, the median AIC at follow-up in the PT group was significantly larger than the median AIC at implantation in the control group (70° vs 58°, respectively; P = 0.014). CONCLUSION: The HeartMate II LVAD IC position contributes to the development of PT. Regular monitoring of cannula position may help identify patients at risk for this problem.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。