Characterization of an MR-compatible motion platform for quality assurance of motion-compensated treatments on the 1.5 T MR-linac

对1.5T磁共振直线加速器上运动补偿治疗质量保证的磁共振兼容运动平台进行特性分析

阅读:1

Abstract

BACKGROUND: Novel motion-compensated treatment techniques on the MR-linac can address adverse intra-fraction motion effects. These techniques involve beam gating or intra-fraction adaptations of the treatment plan based on real-time magnetic resonance imaging (MRI) performed during treatment. For quality assurance (QA) of these workflows, a multi-purpose motion platform is desirable. This platform should accommodate various phantoms, enabling multiple QA workflows. PURPOSE: This study aims to evaluate the new IBA QUASAR Motion MR Platform for use in the 1.5 T MR-linac. METHODS: The motion platform was assessed for several magnetic resonance (MR) characteristics, including spurious noise generation and B0&B1 homogeneity. In addition, the motion platform's motion accuracy and beam attenuation were assessed. An application was shown with a ScandiDos Delta4 Phantom+ MR demonstrating patient-specific plan QA of gated treatments using time-resolved dosimetry that includes motion based on a patient's respiratory motion trace. RESULTS: All MR characterization measurements were within the set tolerances for MRI QA. The motion platform motion accuracy showed excellent agreement with the reference, with a standard deviation of the amplitude of 0.01  mm (20 kg load) for the motor's self-estimated positions and 0.22 mm (no load) for the images acquired with the electronic portal imager. Beam attenuation was found to be 11.8%. The combination of the motion platform and Delta4 demonstrated motion-included dosimetry at high temporal and spatial resolutions. Motion influenced the measured dose in non-gated treatments by up to -20.1%, while gated deliveries showed differences of up to -1.7% for selected diodes. CONCLUSION: The motion platform was found to be usable in a 1.5 T magnetic field, and for all MR characterization experiments, no influence from the motion platform was observed. This motion platform enables to perform motion-included QA, with a measurement device of choice.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。