Commissioning of a Commercial Secondary Dose Check Software and Clinical Implementation for the Magnetic Resonance-guided Linear Accelerator Adaptive Workflow

商用二级剂量检查软件的调试及磁共振引导直线加速器自适应工作流程的临床实施

阅读:2

Abstract

PURPOSE: The purpose of this study was to report the commissioning the secondary dose calculation software ThinkQA (TQA) for an magnetic resonance-guided linear accelerator (MR-linac). METHODS: The Medical Physics Practice Guideline 5.a. (MPPG5a) tests, and dose in inhomogeneities, beam profiles, and depth dose curves were calculated and compared between Monaco and TQA. Five intensity modulated radiotherapy (IMRT) plans (anal, abdominal, head and neck, prostate, and lung), based on TG-244 guidelines were evaluated varying the gamma criteria. Furthermore, the initial and adapted plans for the first session for 17 patients in different anatomical regions were calculated in TQA using different gamma criteria. For five patients, six measurements were made at different fractions using ArcCheck and compared with TQA. RESULTS: The majority of tests met the tolerances defined in the MPPG5a with the exception of dose profiles (>10%), and large multileaf collimator-shaped fields with extensive blocking (>2%). For the IMRT plans, tight criteria such as 2%/2 mm may not be suitable for all scenarios. Thus, we adopt a reasonable 3%/2 mm without compromising the quality of the plan that included significant high-to-low-density interfaces. It is observed that, the values obtained for clinical cases are in the range from 94.6% to 99.8% (TQA), 97.0% to 99.6% (ArcCheck), except in a prostate case with 87.8% (TQA) and 99.3% (ArcCheck). CONCLUSION: We commissioned TQA as a secondary dose calculation for MR-linac and we introduced it clinically for adaptive treatment workflow using 3%/2 mm with 95% as tolerance limit and 90% as action limit.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。