Assessing the Organ Dose in Diagnostic Imaging with Digital Tomosynthesis System Using TLD100H Dosimeters

利用TLD100H剂量计评估数字断层合成系统中诊断成像的器官剂量

阅读:1

Abstract

BACKGROUND: Digital tomosynthesis (DTS) is an advanced imaging modality that enhances diagnostic accuracy by offering three-dimensional visualization from two-dimensional projections, which is particularly beneficial in breast and lung imaging. However, this increased imaging capability raises concerns about patient exposure to ionizing radiation. METHODS: This study explores the energy and angular dependence of thermoluminescent dosimeters (TLDs), specifically TLD100H, to improve the accuracy of organ dose assessment during DTS. Using a comprehensive experimental approach, organ doses were measured in both DTS and traditional RX modes. RESULTS: The results showed lung doses of approximately 3.21 mGy for the left lung and 3.32 mGy for the right lung during DTS, aligning with the existing literature. In contrast, the RX mode yielded significantly lower lung doses of 0.33 mGy. The heart dose during DTS was measured at 2.81 mGy, corroborating findings from similar studies. CONCLUSIONS: These results reinforce the reliability of TLD100H dosimetry in assessing radiation exposure and highlight the need for optimizing imaging protocols to minimize doses. Overall, this study contributes to the ongoing dialogue on enhancing patient safety in diagnostic imaging and advocates for collaboration among medical physicists, radiologists, and technologists to establish best practices.