Dosimetric commissioning of a high-resolution CMOS 2D detector array for patient-specific QA of single-isocenter multi-target VMAT stereotactic radiosurgery

用于单中心多靶点VMAT立体定向放射外科手术患者特定QA的高分辨率CMOS二维探测器阵列的剂量学调试

阅读:1

Abstract

Stereotactic radiosurgery (SRS) using the single-isocenter-multiple-target (SIMT) technique by volumetric modulated arc therapy is increasingly popular for treating multiple brain metastases. However, the complex nature of SIMT SRS necessitates rigorous patient-specific quality assurance (PSQA). This study presents a multi-institutional dosimetric commissioning of a high-resolution complementary metal oxide semiconductor (CMOS) 2D detector array, the myQA SRS device for SIMT SRS PSQA. Basic dosimetric properties such as dose-rate, field-size, energy and angular dependencies were characterized for the CMOS detectors. Additionally, gamma index analyses were performed between the measured dose and the films for nine simulated and clinical plans. The results showed that the CMOS detector was dose-rate, field-size, energy and beam-angle dependent. Specific to SIMT SRS, angular dependence on gantry rotations was invariant to couch rotations but was sensitive to off-isocenter distances. With appropriate dose calibration and angular corrections, myQA SRS showed a high dosimetric correlation with films. The average gamma index pass rates were 99.9 ± 0.03% and 99.2 ± 1.1% at 3%/2 mm/10%thr(global) and 1 mm/1%/10%thr(local) criteria, respectively. The average dose difference between myQA SRS and films was 0.4 ± 1.3%. In conclusion, the CMOS 2D detector array has demonstrated its potential as a reliable tool for PSQA for SIMT SRS. The excellent dosimetric agreement with the films was consistent in multiple institutions, further validating the dosimetric accuracy and reproducibility. It provides a timely alternative to film dosimetry for commissioning and quality assurance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。