Linearity and bias of proton density fat fraction across the full dynamic range of 0-100%: a multiplatform, multivendor phantom study using 1.5T and 3T MRI at two sites

在0-100%的全动态范围内,质子密度脂肪分数的线性度和偏差:一项在两个中心使用1.5T和3T磁共振成像仪的多平台、多厂商体模研究

阅读:1

Abstract

OBJECTIVE: Performance assessments of quantitative determinations of proton density fat fraction (PDFF) have largely focused on the range between 0 and 50%. We evaluate PDFF in a two-site phantom study across the full 0-100% PDFF range. MATERIALS AND METHODS: We used commercially available 3D chemical-shift-encoded water-fat MRI sequences from three MRI system vendors at 1.5T and 3T and conducted the study across two sites. A spherical phantom housing 18 vials spanning the full 0-100% PDFF range was used. Data at each site were acquired using default parameters to determine same-day and different-day intra-scanner repeatability, and inter-system and inter-site reproducibility, in addition to linear regression between reference and measured PDFF values. RESULTS: Across all systems, results demonstrated strong linearity and minimal bias. For 1.5T systems, a pooled slope of 0.99 with a 95% confidence interval (CI) of 0.981-0.997 and a pooled intercept of 0.61% PDFF with a 95% CI of 0.17-1.04 were obtained. Results for pooled 3T data included a slope of 1.00 (95% CI 0.995-1.005) and an intercept of 0.69% PDFF (95% CI 0.39-0.97). Inter-site and inter-system reproducibility coefficients ranged from 2.9 to 6.2 (in units of PDFF), while intra-scanner same-day and different-day repeatability ranged from 0.6 to 7.8. DISCUSSION: PDFF across the 0-100% range can be reliably estimated using current commercial offerings at 1.5T and 3T.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。