Dosimetric impact of calibration coefficients determined using linear accelerator photon and electron beams for ionization chamber in an on-site dosimetry audit

在现场剂量学审核中,利用直线加速器光子束和电子束测定电离室校准系数的剂量学影响

阅读:1

Abstract

This study aimed to clarify the dosimetric impact of calibration beam quality for calibration coefficients of the absorbed dose to water for an ionization chamber in an on-site dosimetry audit. Institution-measured doses of 200 photon and 184 electron beams were compared with the measured dose using one year data before and after the calibration of the ionization chamber used. For photon and electron reference dosimetry, the agreements of the institution-measured dose against two measured doses in this audit were evaluated using the calibration coefficients determined using 60Co (${N}_{D,\mathrm{w},{}^{60}\mathrm{Co}}$) and linear accelerator (linac) (${N}_{D,\mathrm{w},Q}$) beams. For electron reference dosimetry, the agreement of two institution-measured doses against the measured dose was evaluated using${N}_{D,\mathrm{w},Q}$. Institution-measured doses were evaluated using direct- and cross-calibration coefficients. For photon reference dosimetry, the mean differences and standard deviation (SD) of institution-measured dose against the measured dose using ${N}_{D,\mathrm{w},{}^{60}\mathrm{Co}}$ and ${N}_{D,\mathrm{w},Q}$ were -0.1% ± 0.4% and -0.3% ± 0.4%, respectively. For electron reference dosimetry, the mean differences and SD of institution-measured dose using the direct-calibration coefficient against the measured dose using ${N}_{D,\mathrm{w},{}^{60}\mathrm{Co}}$ and ${N}_{D,\mathrm{w},Q}$ were 1.3% ± 0.8% and 0.8% ± 0.8%, respectively. Further, the mean differences and SD of institution-measured dose using the cross-calibration coefficient against the measured dose using ${N}_{D,\mathrm{w},Q}$ were -0.1% ± 0.6%. For photon beams, the dosimetric impact of introducing calibration coefficients determined using linac beams was small. For electron beams, it was larger, and the measured dose using ${N}_{D,\mathrm{w},Q}$ was most consistent with the institution-measured dose, which was evaluated using a cross-calibration coefficient.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。