Real-time tumor-tracking radiotherapy with SyncTraX for primary liver tumors requiring isocenter shift†

SyncTraX实时肿瘤追踪放射治疗技术适用于需要等中心偏移的原发性肝肿瘤†

阅读:1

Abstract

The SyncTraX series enables real-time tumor-tracking radiotherapy through the real-time recognition of a fiducial marker using fluoroscopic images. In this system, the isocenter should be located within approximately 5-7.5 cm from the marker, depending on the version, owing to the limited field of view. If the marker is placed away from the tumor, the isocenter should be shifted toward the marker. This study aimed to investigate stereotactic body radiotherapy (SBRT) outcomes of primary liver tumors treated with SyncTraX in cases where the isocenter was shifted marginally or outside the planning target volume (PTV). Twelve patients with 13 liver tumors were included in the analysis. Their isocenter was shifted toward the marker and was placed marginally or outside the PTV. The prescribed doses were generally 40 Gy in four fractions or 48 Gy in eight fractions. The overall survival (OS) and local control (LC) rates were calculated using the Kaplan-Meier method. All patients completed the scheduled SBRT. The median distance between the fiducial marker and PTV centroid was 56.0 (interquartile range [IQR]: 52.7-66.7) mm. By shifting the isocenter toward the marker, the median distance between the marker and isocenter decreased to 34.0 (IQR: 33.4-39.7) mm. With a median follow-up period of 25.3 (range: 6.9-70.0) months, the 2-year OS and LC rates were 100.0% (95% confidence interval: 100-100). An isocenter shift makes SBRT with SyncTraX feasible in cases where the fiducial marker is distant from the tumor.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。