Development, Construction, and Evaluation of an Alternative Dosimetry Phantom for Computed Tomography

计算机断层扫描替代剂量测定体模的开发、构建和评估

阅读:1

Abstract

This article aims to present the development, construction, and evaluation of an alternative computed tomography dose index (CTDI) phantom. Epoxy resin was mixed with an iodine-based contrast agent to produce radiological characteristics resembling polymethyl methacrylate (PMMA) as a standard CTDI phantom. As a preliminary study, testing was carried out using computed tomography images (80 and 120 kVp) on 12 variations of epoxy-iodine resin mixtures to obtain relative electron density (ρ(e)) values and effective atomic numbers (Z(eff)) of the samples. The alternative CTDI phantoms were then constructed with a resin-iodine mixture using iodine concentrations that yield on closest ρ(e) and Z(eff) values to those of PMMA. The evaluation was carried out by comparing dose measurement results at various energies between the alternative phantom and the International Electrotechnical Commission-standard CTDI phantom. At a concentration of 0.46%, the epoxy resin has ρ(e) and Z(eff) with a deviation against PMMA of 0.12% and 1.58%, respectively, so that composition was chosen for the alternative CTDI phantom construction. The average dose discrepancy values were 5% and 1%, respectively, for the head and body phantoms in the tested tube voltages of 80 kVp, 100 kVp, 120 kVp, and 135 kVp. The Student's t-test result between the alternative and the standard phantoms also showed P < 0.05, indicating the comparability of the alternative CTDI phantom with the standard CTDI phantom.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。