Overcoming Problems Caused by Offset Distance of Multiple Targets in Single-isocenter Volumetric Modulated Arc Therapy Planning for Stereotactic Radiosurgery

克服单中心容积调强弧形治疗立体定向放射外科计划中多靶点偏移距离引起的问题

阅读:1

Abstract

PURPOSE: The purpose of the study is to investigate the impact of large target offset distances on the dose distribution and gamma passing rate (GPR) in single-isocenter multiple-target stereotactic radiosurgery (SIMT SRS) using volumetric modulated arc therapy (VMAT) with a flattening filter-free (FFF) beam from a linear accelerator. METHODS: Two targets with a diameter of 1 cm were offset by "±2, ±4, and ±6 cm from the isocenter in a verification phantom for head SRS (20 Gy/fr). The VMAT plans were created using collimator angles that ensured the two targets did not share a leaf pair from the multi-leaf collimator. To evaluate the low-dose spread intermediate dose spill (R(50%)), GPRs were measured with a criterion of 3%/2 mm using an electronic portal imaging device and evaluated using monitor unit (MU), modulation complexity score for VMAT (MCS(v)), and leaf travel (LT) parameters. RESULTS: For offsets of 2, 4, and 6 cm, the respective parameters were: R(50%), 4.75 ± 0.36, 5.13 ± 0.36, and 5.11 ± 0.33; GPR, 95.01%, 93.82%, and 90.67%; MU, 5893 ± 186, 5825 ± 286, and 5810 ± 396; MCS(v), 0.24, 0.16, and 0.13; and LT, 189.21 ± 36.04, 327.69 ± 67.01, and 430.39 ± 114.34 mm. There was a spread in the low-dose region from offsets of ≥4 cm and the GPR negatively correlated with LT (r = -0.762). There was minimal correlation between GPR and MU or MCS(v). CONCLUSIONS: In SIMT SRS VMAT plans with an FFF beam from a linear accelerator, target offsets of <4 cm from the isocenter can minimize the volume of the low-dose region receiving 10 Gy or more. During treatment planning, it is important to choose gantry, couch, and collimator angles that minimize LT and thereby improve the GPR.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。