Right ventricular lead perforation with iatrogenic injury to an intercostal artery causing haemothorax after pacemaker implant

起搏器植入术后,右心室导线穿孔并伴有医源性肋间动脉损伤,导致血胸

阅读:2

Abstract

A woman in her 80s experienced a life-threatening complication of pacemaker implant consisting of subacute right ventricular lead perforation causing iatrogenic injury to an intercostal artery, resulting in a large haemothorax. A CT scan confirmed active bleeding from the fourth intercostal artery. The patient underwent cardiothoracic surgery via a median sternotomy approach, during which the source of the bleeding was sealed, a new epicardial lead was positioned, and the original lead was extracted. This case emphasises the potentially severe consequences of pacemaker lead perforation and secondary injury to adjacent structures. It underscores the importance of early recognition and timely intervention, preferably in a tertiary specialist unit equipped for cardiothoracic surgery and confirms the value of pacemaker interrogation and CT scans for diagnosis.