Comparison of reconstructed prompt gamma emissions using maximum likelihood estimation and origin ensemble algorithms for a Compton camera system tailored to proton range monitoring

针对质子射程监测,对康普顿相机系统采用最大似然估计和原点集合算法重建的瞬发伽马射线发射进行比较

阅读:1

Abstract

Compton-based prompt gamma (PG) imaging is being investigated by several groups as a potential solution for in vivo range monitoring in proton therapy. The performance of this technique depends on the detector system as well as the ability of the reconstruction method to obtain good spatial resolution to establish a quantitative correlation between the PG emission and the proton beam range in the patient. To evaluate the feasibility of PG imaging for range monitoring, we quantitatively evaluated the emission distributions reconstructed by a Maximum Likelihood Expectation Maximization (MLEM) and a Stochastic Origin Ensemble (SOE) algorithm. To this end, we exploit experimental and Monte Carlo (MC) simulation data acquired with the Polaris-J Compton Camera (CC) prototype. The differences between the proton beam range (R(D)) defined as the 80% distal dose fall-off and the PG range (R(PG)), obtained by fitting the distal end of the reconstructed profile with a sigmoid function, were quantified. A comparable performance of both reconstruction algorithms was found. For both experimental and simulated irradiation scenarios, the correlation between R(D) and R(PG) was within 5 mm. These values were consistent with the ground truth distance (R(D)-R(PGg)≈ 3 mm) calculated by using the expected PG emission available from MC simulation. Furthermore, shifts of 3 mm in the proton beam range were resolved with the MLEM algorithm by calculating the relative difference between the R(PG) for each reconstructed profile. In non-homogeneous targets, the spatial changes in the PG emission due to the different materials could not be fully resolved from the reconstructed profiles; however, the fall-off region still resembled the ground truth emission. For this scenario, the PG correlation (R(D)-R(PG)) varied from 0.1 mm to 4 mm, which is close to the ground truth correlation (3 mm). This work provides a framework for the evaluation of the range monitoring capabilities of a CC device for PG imaging. The two investigated image reconstruction algorithms showed a comparable and consistent performance for homogeneous and heterogeneous targets.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。