Early migration of a medially stabilized total knee arthroplasty : a radiostereometric analysis study up to two years

内侧稳定型全膝关节置换术早期移位:一项长达两年的放射立体分析研究

阅读:2

Abstract

AIMS: Medial pivot (MP) total knee arthroplasties (TKAs) were designed to mimic native knee kinematics with their deep medial congruent fitting of the tibia to the femur almost like a ball-on-socket, and a flat lateral part. GMK Sphere is a novel MP implant. Our primary aim was to study the migration pattern of the tibial tray of this TKA. METHODS: A total of 31 patients were recruited to this single-group radiostereometric analysis (RSA) study and received a medial pivot GMK Sphere TKA. The distributions of male patients versus female patients and right versus left knees were 21:10 and 17:14, respectively. Mean BMI was 29 kg/m(2) (95% confidence interval (CI) 27 to 30) and mean age at surgery was 63 years (95% CI 61 to 66). Maximum total point motions (MTPMs), medial, proximal, and anterior translations and transversal, internal, and varus rotations were calculated at three, 12, and 24 months. Patient-reported outcome measure data were also retrieved. RESULTS: MTPMs at three, 12, and 24 months were 1.0 mm (95% CI 0.8 to 1.2), 1.3 mm (95% CI 0.9 to 1.7), and 1.4 mm (0.8 to 2.0), respectively. The Forgotten Joint Score was 79 (95% CI 39 to 95) and Knee Injury and Osteoarthritis Outcome Score obtained at two years was 94 (95% CI 81 to 100), 86 (95% CI 75 to 93), 94 (95% CI 88 to 100), 69 (95% CI 48 to 88), and 81 (95% CI59 to 100) for Pain, Symptoms, Activities of Daily Living, Sport & Recreation, and Quality of Life, respectively. CONCLUSION: In conclusion, we found that the mean increase in MTPM was lower than 0.2 mm between 12 and 24 months and thus apparently stable. Yet the GMK Sphere had higher migration at one and two years than anticipated. Based on current RSA data, we therefore cannot conclude on the long-term performance of the implant, pending further assessment. Cite this article: Bone Jt Open 2021;2(9):737-744.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。