Performance of a ⁶²Zn/⁶²Cu microgenerator in kit-based synthesis and delivery of [⁶²Cu]Cu-ETS for PET perfusion imaging

⁶²Zn/⁶²Cu微型发生器在基于试剂盒的[⁶²Cu]Cu-ETS合成和递送中的性能,用于PET灌注成像

阅读:1

Abstract

The performance of a commercially produced (62)Zn/(62)Cu microgenerator system, and an associated kit-based radiopharmaceutical synthesis method, was evaluated for clinical site production of [(62)Cu]Cu-ETS (ethylglyoxal bis(thiosemicarbazonato)copper(II)), an investigational agent for PET perfusion imaging. Using 37 generators, containing 1.84±0.23 GBq (62)Zn at 9:00 AM on the day of clinical use, a total of 45 patient doses of [(62)Cu]Cu-ETS (672±172 MBq) were delivered without difficulty. (62)Cu elution yields were high (approximately 90%), accompanied by extremely low (62)Zn breakthrough (<0.001%). Radiopharmaceutical preparation, from the start-of-elution to time-of-injection, consumed less than five minutes. The (62)Zn/(62)Cu microgenerator was a dependable source of short-lived positron-emitting (62)Cu, and the kit-based synthesis proved to be rapid, robust, and highly reliable for "on-demand" delivery of [(62)Cu]Cu-ETS for PET perfusion imaging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。