A method to prevail false positive responses due to excess cations and viscous nature of Radiopharmaceuticals in Limulus Amebocyte Lysate Gel Clot test

一种克服鲎变形细胞溶解物凝胶凝块试验中因过量阳离子和放射性药物粘稠性引起的假阳性反应的方法

阅读:1

Abstract

OBJECTIVES: Bacterial endotoxin test (BET) for detection and quantification of endotoxin in radiopharmaceuticals (RPs), used for therapy or diagnosis, is prerequisite to administration in patients. Out of the two established methods used for this purpose (Kinetic Chromogenic Assay: KCM and Gel Clot Bacterial Endotoxin Test: GC-BET), GC-BET is recommended by pharmacopeias to evaluate the interferences exhibited during the assay due to presence of various ingredients in samples. In the present study, the influence of excess of cations in [(177)Lu]Lu-DOTATATE, used for Peptide Receptor Radionuclide Therapy (PRRT), were studied and a protocol to negate the enhancement observed was developed. Additionally, a protocol for carrying out GC-BET for extremely viscous [(131)I]I-Lipiodol was standardized. METHODS: GC-BET was performed for [(177)Lu]Lu-DOTATATE and [(131)I]I-Lipiodol at maximum valid dilution (MVD), using LRW as a diluent. To negate the false positivity observed in case of [(177)Lu]Lu-DOTATATE, various concentrations of calcium chloride (CaCl(2)) were added and evaluated for the reversal of the interference observed initially. To prevail the difficulty in performing GC-BET for [(131)I]I-Lipiodol various modification in the protocols like orbital vortexing at different rpm and time intervals were performed. KCM assays were also performed for studied RPs at MVD. RESULTS: It was observed that at MVD, [(177)Lu]Lu-DOTATATE exhibited false positivity in GC-BET. However, all the individual reagents used in labeling of [(177)Lu]Lu-DOTATATE did not show any false positivity. Finally, performing the assay with an addition of 2mM CaCl(2) (final concentration) nullified the false positivity. Further, intricacy in performing GC-BET for [(131)I]I-Lipiodol due to its viscosity was resolved by orbital vortexing at 3000 rpm for 5 minutes. CONCLUSIONS: Our study proved that false positivity was observed in GC-BET for [(177)Lu]Lu-DOTATATE due to the presence excess M(3+) ions. Further, our study is the first of its kind which demonstrated methods for negating these false positive results by using modified protocol and hypothesizing the reason behind the enhancement. Additionally, ours is the first study which proved that a simple step of vortexing the viscous RPs like [(131)I]I-Lipiodol can resolved the problems encountered during performing GC-BET due to viscosity of RPs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。