A Randomized Crossover Trial of a Pressure Relief Technology (SensAwake™) in Continuous Positive Airway Pressure to Treat Obstructive Sleep Apnea

一项关于压力释放技术(SensAwake™)在持续气道正压通气治疗阻塞性睡眠呼吸暂停的随机交叉试验

阅读:3

Abstract

OBJECTIVES/BACKGROUND: Improving adherence to CPAP devices is crucial to reduce the long-term morbidity associated with OSA. SensAwake is a unique pressure relief technology that aims to promptly reduce the pressure upon sensing irregular respiration indicative of wakefulness. The purpose of this study was to compare adherence and sleep-quality outcomes in patients treated by CPAP with and without SensAwake technology. METHODS: Participants with moderate-to-severe OSA were randomized to use CPAP devices with or without SensAwake (4 weeks) before crossing over. RESULTS: Sixty-five patients completed both arms of the trial. There were no statistically significant differences in CPAP adherence with or without SensAwake over the study period (SensAwake ON 272.67 ± 17.06 versus SensAwake OFF 289.09 ± 15.24; p = 0.180). SensAwake reported a significantly lower system leak, 90th percentile leak, and time spent with excessive (>60 L/min) leak. Subgroup analysis suggested a trend towards improved adherence in patients with moderate-to-severe insomnia when using SensAwake. CONCLUSIONS: Using SensAwake incurred benefit in terms of reduced leaks; however, SensAwake did not improve CPAP adherence or objective sleep quality. Further studies should investigate the accuracy of observed trends towards increased adherence using SensAwake among patients with OSA and insomnia.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。