Background
Synthetic cannabinoids are touted as legal alternatives to cannabis, at least when first released, and routine urine cannabinoid screening
Conclusions
This high throughput immunoassay exhibited good diagnostic efficiency and documented that the NMS JWH-018 direct ELISA is a viable method for screening synthetic cannabinoids in urine targeting the JWH-018 N-(5-hydroxypentyl) and related analytes. Optimal performance was achieved with a matrix-matched 5 mcg/L urine cutoff.
Methods
The NMS ELISA kit targeting the JWH-018 N-(5-hydroxypentyl) metabolite was used to screen 2492 urine samples with 5 and 10 mcg/L cutoffs. A fully validated liquid chromatography-tandem mass spectrometry method for 29 synthetic cannabinoids markers confirmed all presumptive positive and negative
Results
The immunoassay was linear from 1 to 500 mcg/L with intraplate and interplate imprecision of ≤8.2% and <14.0%, respectively. No interferences were present from 93 common drugs of abuse, metabolites, coadministered drugs, over-the-counter medications, or structurally similar compounds, and 19 of 73 individual synthetic cannabinoids (26%) exhibited moderate to high cross-reactivity to JWH-018 N-(5-hydroxypentyl) metabolite. Sensitivity, specificity, and efficiency results were 83.7%, 99.4%, and 97.6%, as well as 71.6%, 99.7%, and 96.4% with the 5 and 10 mcg/L urine cutoffs, respectively. Conclusions: This high throughput immunoassay exhibited good diagnostic efficiency and documented that the NMS JWH-018 direct ELISA is a viable method for screening synthetic cannabinoids in urine targeting the JWH-018 N-(5-hydroxypentyl) and related analytes. Optimal performance was achieved with a matrix-matched 5 mcg/L urine cutoff.