Distribution and Repeatability of Corneal Astigmatism Measurements (Magnitude and Axis) Evaluated With Color Light Emitting Diode Reflection Topography

利用彩色发光二极管反射地形图评估角膜散光测量(度数和轴位)的分布和重复性

阅读:1

Abstract

PURPOSE: To evaluate and investigate the distribution and repeatability of anterior corneal surface astigmatism measurements (axis and magnitude) using a novel corneal topographer. METHODS: Anterior corneal surface astigmatism was investigated in a total of 195 eyes using a novel multicolored spot reflection topographer (Cassini; i-Optics). Two patient groups were studied, a younger-age group A and an older-age group B. Three consecutive acquisitions were obtained from each eye. The repeatability of measurement was assessed using Bland-Altman plot analysis and is reported as the coefficient of repeatability. RESULTS: Group A (average age 34.3 years) had on average with-the-rule astigmatism, whereas the older-age group B (average age 72.3 years) had on average against-the-rule astigmatism. Average astigmatism magnitude measurement repeatability in group A was 0.4 diopters (D) and in group B 0.4 D. Average astigmatism axis measurement repeatability in group A was 5.4 degrees and in group B 5.5 degrees. The axis measurement repeatability improved with increasing magnitude of astigmatism: in the subgroups with astigmatism between 3.0 and 6.0 D, the axis repeatability was 1.4 degrees (group A) and 1.2 degrees (group B), whereas in the subgroups with astigmatism larger than 6.0 D, the repeatability was 1.1 and 0.6 degrees, respectively. CONCLUSIONS: This novel corneal topography device seems to offer high precision in reporting corneal astigmatism. This study reaffirms the established trend of a corneal astigmatism shift from an average "with-the-rule" to "against-the-rule" with aging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。