Data of CT bow tie filter profiles from three modern CT scanners

来自三台现代CT扫描仪的CT蝶形滤波器轮廓数据

阅读:1

Abstract

As one of the key hardware components in Computed Tomography (CT) scanners, a bowtie filter reduces unnecessary radiation dose to the peripheries of a patient and equalizes radiation signal to the detector. Knowledge of the exact profiles from different bowtie filters are critical to model the imaging process of CT scanners and to estimate patient dose. However, bowtie filter profiles remain proprietary to most CT vendors. The data of bowtie profiles reported here were directly measured using a solid-state linear-array detector from three modern CT scanners (GE Revolution, Philips iQon, and Toshiba Aquilion ONE Vision. The detailed method, including associated geometrical calibration and data validation, was previously published. This data article is mainly to present the updated bowtie profile data measured after our previous publication.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。