High-precision optical measurements of 13C/12C isotope ratios in organic compounds at natural abundance

利用高精度光学方法测量天然丰度有机化合物中<sup>13</sup>C/<sup>12</sup>C同位素比值。

阅读:1

Abstract

A continuous-flow cavity ring-down spectroscopy (CRDS) system integrating a chromatographic separation technique, a catalytic combustor, and an isotopic (13)C/(12)C optical analyzer is described for the isotopic analysis of a mixture of organic compounds. A demonstration of its potential is made for the geochemically important class of short-chain hydrocarbons. The system proved to be linear over a 3-fold injection volume dynamic range with an average precision of 0.95 per thousand and 0.67 per thousand for ethane and propane, respectively. The calibrated accuracy for methane, ethane, and propane is within 3 per thousand of the values determined using isotope ratio mass spectrometry (IRMS), which is the current method of choice for compound-specific isotope analysis. With anticipated improvements, the low-cost, portable, and easy-to-use CRDS-based instrumental setup is poised to evolve into a credible challenge to the high-cost and complex IRMS-based technique.