A Novel Classification Technique of Arteriovenous Fistula Stenosis Evaluation Using Bilateral PPG Analysis

一种基于双侧光电容积脉搏波描记法(PPG)分析的动静脉瘘狭窄评估新分类技术

阅读:1

Abstract

The most common treatment for end-stage renal disease (ESRD) patients is the hemodialysis (HD). For this kind of treatment, the functional vascular access that called arteriovenous fistula (AVF) is done by surgery to connect the vein and artery. Stenosis is considered the major cause of dysfunction of AVF. In this study, a noninvasive approach based on asynchronous analysis of bilateral photoplethysmography (PPG) with error correcting output coding support vector machine one versus rest (ESVM-OVR) for the degree of stenosis (DOS) evaluation is proposed. An artificial neural network (ANN) classifier is also applied to compare the performance with the proposed system. The testing data has been collected from 22 patients at the right and left thumb of the hand. The experimental results indicated that the proposed system could provide positive predictive value (PPV) reaching 91.67% and had higher noise tolerance. The system has the potential for providing diagnostic assistance in a wearable device for evaluation of AVF stenosis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。