The Design, Fabrication, and Testing of an Electromagnetic Micropump with a Matrix-Patterned Magnetic Polymer Composite Actuator Membrane

采用矩阵图案化磁性聚合物复合材料致动器膜的电磁微泵的设计、制造和测试

阅读:1

Abstract

A valveless electromagnetic (EM) micropump with a matrix-patterned magnetic polymer composite actuator membrane structure was successfully designed and fabricated. The composite membrane structure is made of polydemethylsiloxane (PDMS) that is mixed with magnetic particles and patterned in matrix blocks. The matrix magnetic composite membrane was fabricated using a soft lithography process and expected to have a compact structure having sufficient magnetic force for membrane deformation and maintained membrane flexibility. The magnetic membrane was integrated with the microfluidic system and functionally tested. The experimental results show that a magnetic composite actuator membrane containing of 6% NdFeB is capable of producing a maximum membrane deflection up to 12.87 µm. The functionality test of the EM actuator for fluid pumping resulted in an extremely low sample injection flow rate of approximately 6.523 nL/min. It was also concluded that there is a correlation between the matrix structure of the actuator membrane and the fluid pumping flow rate. The injection flow rate of the EM micropump can be controlled by adjusting the input power supplied to the EM coil, and this is believed to improve the injection accuracy of the drug dosage and have potential in improving the proficiency of the existing drug delivery system.