Modeling of the Electrochemical Motion Sensor Conversion Factor at High Frequencies

高频下电化学运动传感器转换因子的建模

阅读:1

Abstract

The conversion factor of the electrochemical motion sensors at low frequencies is usually quite high. At the same time, it decreases significantly with the increase in frequency. Thus, increasing the conversion factor for high frequencies is essential for practical use. In this work, the theoretical model that allows establishing the basic laws governing the conversion of high-frequency signals in an electrochemical cell has been suggested. The approach was based on the fact that in the case of high frequencies, the diffusion length is less than the distance between the electrodes and the thickness of the channel and it is enough to consider the transformation of the fluid motion into electrical current only near the cathodes. It was found that the signal output current can be represented as the sum of the term which is proportional to the steady-state concentration gradient along the surface on which the cathode is located, and the term proportional to the concentration gradient normal to the surface. Both first and second terms and the total signal current have been calculated for a particular case of a four-electrode planar system. The practical conclusion is that the high frequency conversion factor increases with the interelectrode distance and the channel width decreases compared to the cathode dimension.