BACKGROUND: This study aimed to investigate the relationship between 3-dimensional (3D) alignment and postoperative bone mineral density (BMD) changes with Accolade II tapered wedge stems, which have a different proximal shape from other tapered wedge stems, up to 5 years after primary total hip arthroplasty. METHODS: We retrospectively analyzed the hips of 89 patients who underwent total hip arthroplasty using the second-generation proximally coated cementless stem (Accolade II; Stryker Orthopedics, Mahwah, NJ) at our institution from 2014 to 2018 over a 5-year follow-up. We evaluated the relationship between stem alignment, measured using 3D-templating software, and BMD changes in the 7 Gruen zones and compared the data with those from a former study using other short taper-wedge stems. RESULTS: BMDs in zones 1 and 7 continued to decrease gradually every year after surgery, and BMD in zone 7 showed the largest decrease (21%) from baseline over 5 years. No correlation was found between stem alignment (varus/valgus, flexion/extension, and anteversion/retroversion) and changes in BMD in each zone over 5 years. CONCLUSIONS: Our data showed no correlation between 3D stem alignment and changes in BMD in each Gruen zone over 5 years. This suggests that the Accolade II stem may fit better into any shape of the proximal medullary canal because of its unique characteristics.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。