Experimental fusion of different versions of the total laboratory automation system and improvement of laboratory turnaround time

实验性地融合不同版本的全实验室自动化系统,并提高实验室周转时间

阅读:1

Abstract

BACKGROUND: Use of total laboratory automation (TLA) system has expanded to microbiology and hemostasis and upgraded to second and third generations. We herein report the first successful upgrades and fusion of different versions of the TLA system, thus improving laboratory turnaround time (TAT). METHODS: A 21-day schedule was planned from the time of pre-meeting to installation and clinical sample application. We analyzed the monthly TAT in each menu, distribution of the "out of range for acceptable TAT" samples, and "prolonged time out of acceptable TAT," before and after the upgrade and fusion. RESULTS: We installed and customized hardware, middleware, and software. The one-way CliniLog 2.0 version track, 50.0-m long, was changed to a 23.2-m long one-way 2.0 version and an 18.7-m long two-way 4.0 version. The monthly TAT in the outpatient samples, before and after upgrading the TLA system, were uniformly satisfactory in the chemistry and viral marker menus. However, in the tumor marker menu, the target TAT (98.0% of samples ≤60 minutes) was not satisfied during the familiarization period. There was no significant difference in the proportion of "out of acceptable TAT" samples, before and after the TLA system upgrades (7.4‰ and 8.5‰). However, the mean "prolonged time out of acceptable TAT" in the chemistry samples was significantly shortened to 17.4 (±24.0) minutes after the fusion, from 34.5 (±43.4) minutes. CONCLUSIONS: Despite experimental challenges, a fusion of the TLA system shortened the "prolonged time out of acceptable TAT," indicating a distribution change in overall TAT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。