A flat interface nerve electrode with integrated multiplexer

带有集成多路复用器的平面接口神经电极

阅读:1

Abstract

One of the goals of peripheral nerve cuff electrode development is the design of an electrode capable of selectively activating a specific population of axons in a common nerve trunk. Several designs such as the round spiral electrode or the flat interface nerve electrode (FINE) have shown such ability. However, multiple contact electrodes require many leads, making the implantation difficult and potentially damaging to the nerve. Taking advantage of the flat geometry of the FINE, multiplexers were embedded within the cuff electrode to reduce the number of leads needed to control 32 channels. The circuit was implemented on a polyimide film using off-the-shelf electronic components. The electronic module was surface-mounted directly onto the electrode's flat substrate. Two circuit designs were designed, built, and tested: 1) a single supply design with only two wires but limited to cathodic-first pulse and 2) a dual-supply design requiring three lead wires but an arbitrary stimulation waveform. The electrode design includes 32 contacts in a 1 mm x 8 mm opening. The contact size is 300 microm x 400 microm with access resistance less than 1 k ohm. This electrode is not intended for long-term use, but developed as a feasibility study for future development using low-water-absorption materials such as liquid crystal polymer and an application specific integrated circuit.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。