A Diagnostic Circuit for Crosstalk Detection in Microelectrode Arrays

用于微电极阵列串扰检测的诊断电路

阅读:1

Abstract

Current leakage between channels in microelectrode arrays is a sign of device failure and can lead to shorting of neural signals. The purpose of this project is to detect crosstalk between 32 channels of electrodes. We designed an embedded crosstalk detection system that can stimulate each electrode individually with a constant-current pulse and record voltage transients of the stimulated and adjacent electrodes to generate a matrix of crosstalk values. Charge injection in a phosphate buffered saline solution was used to check the condition of each electrode. A semi-wet condition was then used to determine the percent crosstalk between the channels. The analysis showed that there was minimal crosstalk between the electrodes, except for a known physical defect on the probe. The measurement technique enabled by the electronics circuit has the potential to be used in functional testing and screening of implantable devices.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。