Evaluating the Efficiency of the Cobas 6800 System for BK Virus Detection in Plasma and Urine Samples

评估 Cobas 6800 系统检测血浆和尿液样本中 BK 病毒的效率

阅读:5
作者:Junhyup Song, Sinyoung Kim, Eunmin Kwak, Younhee Park

Abstract

We evaluated the overall performance of the Cobas 6800 BKV test in detecting BK virus (BKV). We examined the imprecision of the Cobas 6800 BKV test and compared the qualitative and quantitative results obtained from the Cobas 6800 BKV test and the Real-Q BKV quantification assay. We assessed 88 plasma and 26 urine samples collected between September and November 2022 from patients with BKV infection using the Real-Q BKV quantitative assay. The lognormal coefficient of variation indicated that the inter-assay precision of the Cobas 6800 BKV test ranged from 13.86 to 33.83%. A strong correlation was observed between the quantitative results obtained using the Cobas 6800 BKV test and the Real-Q BKV quantification assay for plasma samples. The Spearman's rank correlation coefficients (ρ) for plasma, polymerase chain reaction (PCR) media-stabilized urine, and raw urine samples were 0.939, 0.874, and 0.888, respectively. Our analyses suggest that the Cobas 6800 BKV test is suitable for clinical applications owing to the strong correlation between the results obtained using this test and the Real-Q BKV quantification assay in plasma and urine samples. Furthermore, utilizing fresh raw urine samples can be a viable approach for the Cobas 6800 BKV test as it is less labor- and time-intensive.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。