rSO2 Measurement Using NIRS for Lower-Limb Blood Flow Monitoring and Estimation of Safe Balloon Occlusion/Deflation Time in Patients with PAS Who Underwent PBOA during CS

利用近红外光谱(NIRS)测量rSO2,监测下肢血流,并评估接受体外循环下球囊扩张术(PBOA)的肺动脉狭窄(PAS)患者在心脏手术中球囊扩张术(CS)时的安全球囊阻塞/放气时间。

阅读:1

Abstract

We examined the utility of regional oxygen saturation (rSO2) measurement using near-infrared spectroscopy (NIRS) for monitoring lower-limb blood flow and estimate the safe balloon occlusion/deflation time in patients with PAS who underwent prophylactic balloon occlusion of the abdominal artery (PBOA) during cesarean section (CS). During CS, the NIRS probes were positioned on either of the anterior tibial muscles. rSO2 was measured continuously during balloon occlusion/deflation. A cycle consisted of inflating the aortic balloon for 30 min and deflating it for 5 min. The rSO2 before/during balloon occlusion and after 5 min of balloon deflation were evaluated. Sixty-two lower limbs (fifteen women and data from 31 sessions of balloon inflation/deflation) were evaluated. rSO2 during balloon occlusion was significantly lower than rSO2 before balloon occlusion (57.9% ± 9.6% vs. 80.3% ± 6.0%; p < 0.01). There were no significant differences between rSO2 before balloon occlusion and rSO2 after 5 min of balloon deflation (80.3% ± 6.0% vs. 78.7% ± 6.6%; p = 0.07). Postoperatively, the lower limbs showed no ischemic symptoms. NIRS can assess lower-limb rSO2 during PBOA for PAS in real time to determine ischemia severity, duration, and recovery capacity.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。