Characterisation of tissue factor-bearing extracellular vesicles with AFM: comparison of air-tapping-mode AFM and liquid Peak Force AFM

使用 AFM 表征含有组织因子的细胞外囊泡:空气敲击模式 AFM 与液体峰值力 AFM 的比较

阅读:7
作者:Julie Hardij, Francesca Cecchet, Alexandre Berquand, Damien Gheldof, Christian Chatelain, François Mullier, Bernard Chatelain, Jean-Michel Dogné

Conclusion

We recommend liquid PFT mode to analyse vesicles on 5 µg/ml antibody-coated substrates.

Methods

AFM multimode nanoscope III was used for air tapping mode (TM). AFM catalyst was used for liquid Peak Force Tapping (PFT) mode. Vesicles are generated according to Davila et al.'s protocol. Substrates are coated with various concentrations of antibodies, thanks to ethanolamine and glutaraldehyde.

Results

Vesicles were immobilised on antibody-coated surfaces to select tissue factor (TF)-positive vesicles. The size range of vesicles observed in liquid PFT mode is 6-10 times higher than in air mode. This corresponds to the data found in the literature.